『Installation of mass flow meter transmitter』Related information(clamp on meter|electromagnetic meter|venturi meterrotameter|orifice meter|ultrasonic flow meter|mass flow meter|coriolis mass flow meter|coriolis flow meter|magnetic flow meter|magmeter flow meter|magflow flow meter|mag meter flow meter|electromagnetic flow meter|vortex flow meter|turbine flow meter|thermal mass flow meter|thermal flow meter|rotameter flow meter)

1. What are the requirements for the installation of thermal mass flow meters?
The installation requirements for thermal gas mass flow meters include:
1. Whether the flow meter is installed vertically or horizontally with a rolling ridge, it should be kept in a horizontal state.
2. In places where unexpected gas shutdown is not allowed or will inevitably cause significant non replaceable losses, a bypass should be set up.
3. It should be confirmed that there is no strong magnetic field, strong electric field, or strong mechanical vibration near the flow meter.
4. The static grounding of the flow meter should be reliable, but it should not be shared with strong electrical grounding.
5. If the flow meter is installed outdoors, a sunshade should be added to prevent sun exposure.
6. Welding operations are prohibited in explosive environments. 7. The surrounding environment should be confirmed to have no corrosive effect on the aluminum alloy. 8. Ensure that the direction of gas flow is consistent with the direction of the arrow on the flowmeter. 9. There should be very few 10D straight pipe sections in front of the flowmeter, and very few 5D (D is the pipe diameter) straight pipe sections behind it. 10. For environments with special welding requirements, operations should be carried out in accordance with the relevant requirements for large blind spot welding. 11. It is prohibited to install flow meters at low points in the pipeline, otherwise water and other impurities may accumulate.2. How to correctly install a mass flow meter
The following seven steps are required to correctly install a mass flow meter: (1) General points: There should be at least 0.6m distance